69 Gas Analyzer Calibration Error MP4000

Description 

Gas Analyzer Calibration Error 

  • Occurs during AFAM+ PTI test. 
    • After ventilation if the CO2 reading is below 0% or above 2%. 
  • Occurs during normal run with AFAM+ enabled. 
    • With O2 ON, if (CO2+O2) is not within 10% to 30%. 
    • With O2 OFF, if CO2 is above 25%. 
  • Indicates: 
  • Sensor lost the calibration. 
  • Failing sensor. 

Corrective Action 

  • Redo the test. 
  • Replace sensor.