Trouble
shooting |
Possible causes:
- Defective power supply for main controller.
- Defective sensor pulling power supply down.
- Defective main controller.
Accompanied alarms:
- There may be sensor alarms, AL 3XX.
- AL 801 will be activated during FT- or PTI test.
Trouble shooting:
- Check if there are other active (sensor) alarms. Clear these alarms fi rst using their trouble shooting.
- Access the “special menu” by pressing "spanner icon" for more than 3 sec. Scroll down until you see the label “U AirEx” in the display. The value of “U AirEx“ should be in the range between 4.80 and 5.2V DC.
- While displaying “U AirEx“, remove the connector at AirEx potentiometer. − If “U AirEx“ is now inside the correct above range (4.80 – 5.20V DC), the AirEx potentiometer is defective. Replace the AirEx potentiometer − If “U AirEx” is still outside the above range, proceed to next step.
- While displaying “U AirEx“, disconnect “AirEx” from the terminals according to wiring schematics inside in the control cabinet. − If “U AirEx” is now inside the above range the cable for “AirEx” is defective. Mount connector correctly or replace connector cable for “AirEx”. − If the correct voltage is measured at X22 and at X23 then circuit is defective.
Replace the main controller. |